Fluke Biomedical QA-ES III Electrosurgical Analyzer
Electrosurgical Analyzers
$5495from
Category
ESU analyzers and test loads for generator output, leakage and REM/ARM verification.
Electrosurgical Analyzers
$5495from
Electrosurgical Analyzers
$2695from
Electrosurgical Analyzers
$795from
ESU analyzers measure RF output power, current and crest factor into a controlled load, and verify return electrode monitoring. They are annual-inspection equipment for any department maintaining electrosurgical generators.
Measuring RF power requires a precise, non-inductive load that can absorb the generator's full output. The QA-ES III has an internal variable load bank covering 0 to 5115 ohm, which is what allows automated power distribution curves.
For higher-power generators an external load such as the RF303 extends the range.
Return electrode monitoring verification is a safety-critical test that confirms the generator will shut down on a poor patient plate connection. Confirm the analyzer supports the contact quality test method your generators use.
We exercise the load bank through its full range and verify RF measurement against a generator of known output before listing.
For routine power verification, yes. The QA-ES III adds automation and a wider load range.
Yes, though cable adapters vary. Tell us which generators you service and we will confirm the adapters needed.
The resistive elements are durable. We step through the full range as part of testing.
Prices shown are indicative for a single tested unit. Quantity pricing, dealer terms and blanket orders are quoted individually. Send the list and we will respond within one business day.