Category

Incubator & NIBP Analyzers

Infant incubator and radiant warmer analyzers plus dedicated NIBP simulators.

Brands in this category: Fluke Biomedical

Infant incubator analyzers and NIBP simulators cover two small but essential parts of a biomedical department's annual testing programme.

Incubator and radiant warmer testing

The INCU II tests to IEC 60601-2-19, -2-20 and -2-21 using five temperature sensors plus airflow, humidity and sound measurement. Sound level matters more than people expect — noise limits in neonatal care are strictly specified.

Sensor pods are the item to check. All five must be present and functional, and they are model-specific.

NIBP simulation

A BP Pump 2 simulates dynamic blood pressure envelopes from multiple manufacturers, plus static pressure, leak and relief valve testing. Cuffs, T-fittings and hoses are consumable and frequently missing.

The ProSim 8 also includes NIBP simulation, so departments buying one may not need the other.

Common questions

Do I need a dedicated NIBP simulator?

Not if you already have a ProSim 8. A dedicated unit gives faster throughput and more manufacturer envelopes.

Are incubator sensor pods available separately?

Sometimes. Confirm all five are present before purchase, as replacements are not always in stock.

What standards does the INCU II cover?

IEC 60601-2-19 for incubators, -2-20 for transport incubators and -2-21 for radiant warmers.

Ask about incubator & nibp analyzers

Prices shown are indicative for a single tested unit. Quantity pricing, dealer terms and blanket orders are quoted individually. Send the list and we will respond within one business day.